Skip to content

Troy Innovative Instruments Inc

15111 White Rd, Middlefield, Ohio, United States

FEI 3000718985

FDA has 4 inspections on file for Troy Innovative Instruments Inc in Middlefield, Ohio (FEI 3000718985), from 2012 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Jun 30, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.50(a)(2), 21 CFR 820.90(a) and 21 CFR 820.90(b)(2).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2012–2024
1 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Aug 30, 2024
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    4. Supplier oversight

      The type and extent of control to be exercised over services and suppliers was not clearly defined.

    5. Complaints

      Complaint files are not maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    2. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification have not been established.

    2. Supplier oversight

      The type and extent of control to be exercised over services and suppliers was not clearly defined.

    3. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    4. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    Show 3 more citations
    1. Production processes

      Production processes were not developed to ensure that a device conforms to its specifications.

    2. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers and contractors were not evaluated based on their ability to meet specified requirements.

    3. Product rework adverse effects {see also 820.184}

      Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Aug 30, 2024. It asked for a Form 483.

  • Inspection ending Jul 26, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Aug 30, 2024FDA-FOIA-2024-7644Form 483· 1 of 3 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000718985 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000718985. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.