Carolina Liquid Chemistries Corp
510 W Central Ave Ste C, Brea, California, United States
FEI 3000215862
FDA has 8 inspections on file for Carolina Liquid Chemistries Corp in Brea, California (FEI 3000215862), from 2008 to 2026. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
The most recent OAI inspection ended on Nov 9, 2018. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2008–2026
- 1 OAI · 4 VAI · 3 NAI
- Inspections with 483 observations shown
- 3 of 8
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest May 19, 2026
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 7.4.3
Purchased product verification activities not established or implemented
The inspection or other activities necessary for ensuring that purchased product meets specified purchasing requirements were not adequately established and implemented.
- ISO 13485:2016 Clause 8.2.2
Complaint handling procedures lack required elements
Procedures for timely complaint handling do not adequately include requirements and responsibilities for receiving and recording information.
- ISO 13485:2016 Clause 4.1.2
Risk-based approach not applied to QMS processes
A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
Design plans- updated
Design plans were not reviewed and approved as needed as the design and development evolved.
Design input approval documentation
The approval of design input requirements (including the date and signature of the individual approving the requirements) was not documented.
Design review - documentation
The design review results, including identification of the design, the date, and the individual performing the review, were not documented and filed in the design history file.
Design transfer - Lack of or inadequate procedures
Procedures were not implemented to ensure that the device design was correctly transferred into production.
Show 4 more citations
Design output - review and approval
The approval of design output (including the date and signature of the individual approving the output) was not documented.
Design validation - documentation
The results of design validation, including methods and the date, were not documented and filed in the design history file.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities and implementation of corrective and preventive actions.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Compliance with GMP
In vitro diagnostic products are not being manufactured in accordance with the good manufacturing practices set forth in the Quality System Regulation.
DMR device specifications
The device master record does not include or refer to the location of all and device specifications.
Corrective action taken
Records of complaint investigations do not include any corrective action taken.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.
Show 5 more citations
Lot/Control Numbers - multiple unit product
Lot or control numbers on a multiple unit in vitro diagnostic product label and outside container do not permit tracing of the identity of the individual units.
Packaging
Device packaging and/or shipping containers are not designed and constructed well enough to protect the device from damage during processing, storage, handling, and distribution.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been completed.
Legibility
Required records are not legible.
Oral complaints
Complaint handling procedures have not been defined, documented, and completed to ensure that all oral complaints are documented upon receipt.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.
Inspection ending Feb 17, 2026VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 1, 2026. FDAgent holds no copy that FDA posted.
- May 19, 2026FDA-FOIA-2026-5141Form 483· 1 of 4 sitesRecords released
- Date not givenFDA-FOIA-2026-4089Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.130Device packaging1 of 3
- 21 CFR 820.180General requirements1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.198Complaint files1 of 3
Show all 10
- 21 CFR 809.10Labeling for in vitro diagnostic products1 of 3
- 21 CFR 809.20General requirements for manufacturers and producers of in vitro diagnostic products1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000215862 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000215862. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.