Hammill Manufacturing Company
360 Tomahawk Dr, Maumee, Ohio, United States
FEI 3000215103
FDA has 6 inspections on file for Hammill Manufacturing Company in Maumee, Ohio (FEI 3000215103), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Nov 10, 2008. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2008–2025
- 1 OAI · 2 VAI · 3 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to specifications were not established, defined, and implemented.
Specific non-conforming product procedures
Procedures for addressing the documentation, evaluation, and investigation of nonconforming product were not defined and complete.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
DMR device specifications
The device master record does not include or refer to the location of all specifications.
Show 3 more citations
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, implementation of corrective and preventive actions, and dissemination of information about quality problems or nonconforming product to responsible parties.
DMR production process specifications
The device master record does not include or refer to the location of all production and process specifications.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they could be and are being used.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions, implementation of corrective and preventive actions, and dissemination of information about quality problems or nonconforming product to responsible parties.
Document review, approval by designated individual
Documents were not reviewed and approved by the individual designated in document control procedures.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not defined and implemented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Show 9 more citations
Management review participants
Documentation does not show that management with executive responsibility participated in management reviews.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
Document change records, maintained.
Records of changes to documents were not maintained.
DMR production process specifications
The device master record does not include or refer to the location of all production and process specifications.
Nonconforming product evaluation/investigation
The evaluation and investigation of nonconforming product has not been documented.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established.
DMR device specifications
The device master record does not include or refer to the location of all specifications.
Lack of or inadequate organizational structure
The organization structure is not adequate to assure that quality system requirements are fully met.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 820.20Management responsibility2 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.70Production and process controls2 of 4
- 21 CFR 820.75Process validation2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.181Device master record2 of 4
Show all 10
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 820.22Quality audit1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000215103 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000215103. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.