Skip to content

RX Air Industries LLC.

3323 Garden Brook Dr, Farmers Branch, Texas, United States

FEI 3000205843

FDA has 2 inspections on file for RX Air Industries LLC. in Farmers Branch, Texas (FEI 3000205843), from 2011 to 2012. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on May 19, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
2, 2011–2012
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Acceptance activities were not maintained as part of the device history record and adequately documented.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    Show 5 more citations
    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been established.

    4. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been established.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3000205843 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3000205843. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.