Stryker Corporation
5900 Optical Ct, San Jose, California, United States
FEI 2936485
FDA has 7 inspections on file for Stryker Corporation in San Jose, California (FEI 2936485), from 2009 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2009–2019
- 0 OAI · 4 VAI · 3 NAI
- Inspections with 483 observations shown
- 6 of 7
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Oct 30, 2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated where necessary.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely inspected have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Service reports/MDRs/complaints
Service reports that represent MDR reportable events were not automatically considered complaints and processed in accordance with the requirements of 21 CFR 820.198.
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - simulated testing
The design was not validated using production units under actual or simulated use conditions.
Design Validation - Risk analysis not performed/inadequate
Risk analysis was not performed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 30, 2023. Each one asked for a Form 483.
Inspection ending Nov 21, 2019NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 30, 2023FDA-FOIA-2023-9510Form 483Closed
Inspection ending May 25, 2018
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Sep 3, 2021FDA-FOIA-2021-5829Form 483Closed over fees
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 6
- 21 CFR 820.198Complaint files2 of 6
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 806.10Reports of corrections and removals1 of 6
- 21 CFR 820.25Personnel1 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 6
- 21 CFR 820.100Corrective and preventive action1 of 6
Show all 10
- 21 CFR 820.184Device history record1 of 6
- 21 CFR 820.200Servicing1 of 6
Other FEIs recorded as Stryker Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Stryker CorporationFEI 3009390713Minnetonka, Minnesota, United States
Last inspected 2025-06-18NAI
Stryker CorporationFEI 3015967359Portage, Michigan, United States
Last inspected 2023-01-24VAI
Stryker CorporationFEI 1811755Portage, Michigan, United States
Last inspected 2016-01-07NAI
Stryker CorporationFEI 3006011135Portage, Michigan, United States
Last inspected 2010-09-24VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2936485 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2936485. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.