Temrex Corporation
300 Buffalo Ave, Freeport, New York, United States
FEI 2429849
FDA has 6 inspections on file for Temrex Corporation in Freeport, New York (FEI 2429849), from 2010 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.184, 21 CFR 820.40(a) and 21 CFR 820.72(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2010–2025
- 0 OAI · 5 VAI · 1 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Apr 23, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Complaints
Complaint files are not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been established.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained.
Document review, approval documentation
The documentation of approval of documents does not include the document approval date and the signature of the approving official.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Document review, approval by designated individual
Documents were not approved by designated individual(s) prior to issuance .
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria and maintained.
Documented evaluation
The evaluation of potential suppliers and contractors was not documented.
Show 1 more citation
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Equipment control activity documentation
Equipment calibrations and maintenance activities have not been documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Reports and information documentation
The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA.
Equipment control activity documentation
Equipment calibrations and checks have not been documented.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Documentation
Acceptance activities were not documented and maintained as part of the device history record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process controls do not provide for documented instructions, standard operating procedures (SOPs), and methods that define and control the manner of production.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Apr 23, 2025. It asked for a Form 483.
Inspection ending Feb 21, 2025VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 23, 2025FDA-FOIA-2025-3661Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.72Inspection, measuring, and test equipment3 of 6
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 6
- 21 CFR 820.40[Reserved]2 of 6
- 21 CFR 820.50Purchasing controls2 of 6
- 21 CFR 820.100Corrective and preventive action2 of 6
- 21 CFR 820.184Device history record2 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
Show all 10
- 21 CFR 820.86Acceptance status1 of 6
- 21 CFR 820.198Complaint files1 of 6
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2429849 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2429849. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.