Sybaritic, Inc
9220 James Ave S, Bloomington, Minnesota, United States
FEI 2131125
FDA has 11 inspections on file for Sybaritic, Inc in Bloomington, Minnesota (FEI 2131125), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Mar 7, 2018. FDA's compliance record also lists one other action under this FEI: Injunction, . No warning letter or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: n/a and 21 CFR 820.30(i).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 11, 2009–2023
- 5 OAI · 1 VAI · 5 NAI
- Inspections with 483 observations shown
- 4 of 11
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- n/a
Procedural requirements
Deviation from the procedural requirements of a decree of injunction.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- n/a
Procedural requirements
Deviation from the procedural requirements of a decree of injunction.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Investigation procedures
The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.
General
Complaint handling procedures for receiving complaints have not been implemented.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the documentation of design changes before their implementation.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the validation or verification, review, and approval of design changes before their implementation.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.
Lack of or inadequate DHR procedures
Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Lack of procedures, or not maintained
Document control procedures were not implemented.
Training
Employees have not been adequately trained.
Show 3 more citations
Documentation of disposition, justification, signature
There is no documentation of the disposition of nonconforming product, justification for use of nonconforming product, and the signature of the individual authorizing the use of nonconforming product.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 11 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.22Quality audit1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
Show all 11
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.198Complaint files1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- InjunctionCDRH · DevicesFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2131125 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2131125. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.