Skip to content

Sybaritic, Inc

9220 James Ave S, Bloomington, Minnesota, United States

FEI 2131125

FDA has 11 inspections on file for Sybaritic, Inc in Bloomington, Minnesota (FEI 2131125), from 2009 to 2023. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Mar 7, 2018. FDA's compliance record also lists one other action under this FEI: Injunction, . No warning letter or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: n/a and 21 CFR 820.30(i).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
11, 2009–2023
5 OAI · 1 VAI · 5 NAI
Inspections with 483 observations shown
4 of 11
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. n/a

      Procedural requirements

      Deviation from the procedural requirements of a decree of injunction.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  8. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. n/a

      Procedural requirements

      Deviation from the procedural requirements of a decree of injunction.

  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    2. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not implemented.

    3. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not implemented.

    4. General

      Complaint handling procedures for receiving complaints have not been implemented.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures were not followed for the documentation of design changes before their implementation.

  11. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures were not followed for the validation or verification, review, and approval of design changes before their implementation.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not implemented.

    3. Lack of or inadequate DHR procedures

      Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    4. Lack of procedures, or not maintained

      Document control procedures were not implemented.

    5. Training

      Employees have not been adequately trained.

    Show 3 more citations
    1. Documentation of disposition, justification, signature

      There is no documentation of the disposition of nonconforming product, justification for use of nonconforming product, and the signature of the individual authorizing the use of nonconforming product.

    2. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    3. Quality Audit/Reaudit - conducted

      Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 11 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2131125 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2131125. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.