Hampton Medical Devices, LLC
6 Meadow Rdg, Festus, Missouri, United States
FEI 1937445
FDA has 4 inspections on file for Hampton Medical Devices, LLC in Festus, Missouri (FEI 1937445), from 2012 to 2021. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 803.17, 21 CFR 820.198(a) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2012–2021
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration documentation
There is no documentation of calibration dates for inspection and measurement and test equipment .
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented .
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Distribution records
Distribution records do not include or refer to the location of required information.
Examination for accuracy
Labeling was not sufficiently examined by a designated individual for accuracy including before release.
Records, DHR {see also 820.184(e)}
Labels and labeling used for each finished product, lot, or batch, were not sufficiently documented in the DHR.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Remain legible
Labels were not printed so as to remain during processing, storage, handling, distribution, and/or use.
Show 11 more citations
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a method and process have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
- FDCA 501(a)
Medical Device Field Exam
During a field examination of medical device products at your facility the following was observed:
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.120Device labeling1 of 2
Show all 11
- 21 CFR 820.160Distribution1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.184Device history record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1937445 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1937445. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.