B Braun Medical, Inc.
1601 Wallace Dr Ste 150, Carrollton, Texas, United States
FEI 1641965
FDA has 2 inspections on file for B Braun Medical, Inc. in Carrollton, Texas (FEI 1641965), from 2010 to 2012. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 6 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, 2 under 21 CFR Part 806, the device correction and removal reporting rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 806.10(a)(1).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2010–2012
- 2 OAI · 0 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Availability
Required records were not made readily available for review and copying by the FDA and are not backed up when stored in automated data processing systems.
Notification of extended action
No amended notification with required information was submitted within 10 working days of initiating an extension of the correction or removal action.
Show 1 more citation
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Incorrect translation to production specifications
The device design was not correctly translated into production specifications.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 806.10Reports of corrections and removals2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.180General requirements1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1641965 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1641965. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.