Helena Laboratories, Corp.

1530 Lindbergh Dr, Beaumont, Texas, United States

FEI 1618982

FDA has 4 inspections on file for Helena Laboratories, Corp. in Beaumont, Texas (FEI 1618982), from 2010 to 2018. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.70(g)(1), 21 CFR 820.75(a) and 21 CFR 820.100(b).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2010–2018
0 OAI · 4 VAI · 0 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.

    2. Documentation

      Process validation activities and results have not been adequately documented.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been established.

    4. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely inspected and checked have not been adequately established.

    5. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a method and procedure have not been adequately established.

    Show 3 more citations
    1. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    2. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    3. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    2. Design validation - documentation

      The results of design validation, including identification of the design, method(s), and the date, were not adequately documented in the design history file.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    5. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.

    Show 4 more citations
    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Design history file

      The design history file was not established and does not demonstrate that the design was developed following the approved design plan.

    3. Design input - documentation

      Design input requirements were not adequately documented.

    4. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 9

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1618982 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1618982. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.