Medical Supplies Corporation
4002 Dillon St, Baltimore, Maryland, United States
FEI 1119234
FDA has 2 inspections on file for Medical Supplies Corporation in Baltimore, Maryland (FEI 1119234), from 2011 to 2012. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 2 under 21 CFR Part 803, the device adverse event reporting rules, and 2 under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Jul 11, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 803.50(a)(1) and 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2011–2012
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Adverse events--all info not in file
MDR event files do not contain or reference all adverse event information in the possession of the reporting entity, including documentation of the deliberations and decision making process used to determine if an event was or was not reportable.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Show 6 more citations
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Quality policy and objectives
The quality policy and was not established by management with executive responsibility.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate organizational structure
The organizational structure has not been established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.18What are the requirements for establishing and maintaining MDR files or records that apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
Show all 10
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1119234 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1119234. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.