Skip to content

Brant-Wald Surgicals Inc

368 East Tennessee Ave, P. O. Box 4962, Oak Ridge, Tennessee, United States

FEI 1044531

FDA has 3 inspections on file for Brant-Wald Surgicals Inc in Oak Ridge, Tennessee (FEI 1044531), from 2009 to 2011. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Aug 11, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2009–2011
2 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures were not established for monitoring and control of process parameters for validated processes.

    2. Production monitoring and control

      During production, process parameters are not monitored and controlled.

    3. Supplier oversight

      The type and extent of control to be exercised over suppliers and contractors was not clearly defined.

    4. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    5. Maintained

      The acceptance status of product was not clearly identified throughout manufacturing, packaging, and labeling of the product.

    Show 6 more citations
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures were not defined for the control of products that do not conform to specifications.

    2. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers were not documented.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

    5. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not established.

    6. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated, inspected, checked, and maintained were not established.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 3 inspections.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1044531 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1044531. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.