Skip to content

Stat Medical Devices, Inc.

2056 NE 153rd St, North Miami Beach, Florida, United States

FEI 1000525253

FDA has 7 inspections on file for Stat Medical Devices, Inc. in North Miami Beach, Florida (FEI 1000525253), from 2008 to 2025. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Aug 21, 2013. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2008–2025
2 OAI · 4 VAI · 1 NAI
Inspections with 483 observations shown
6 of 7
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Apr 30, 2025
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    2. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    3. Design control - no procedures

      Procedures for design control have not been established.

    4. Training records

      Personnel training is not documented.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by contractors have not been established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    5. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by contractors have not been adequately established.

  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification procedures

      Procedures for verifying that design output meets design input were not implemented.

    2. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers and contractors were not documented and implemented.

    3. Results of investigation

      Records of complaint investigations do not include the results of the investigation.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    5. Documentation

      Process validation activities and results have not been fully documented.

    Show 3 more citations
    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Design verification - documentation

      The design verification results, including identification of the design, methods, the date, and the individual performing the verification, were not documented and filed in the design history file.

    3. Design validation - documentation

      The results of design validation, including identification of the design, methods, the date, and the individual performing the validation, were not documented and filed in the design history file.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Apr 30, 2025. Each one asked for a Form 483.

  • Inspection ending Mar 18, 2025VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Apr 30, 2025FDA-FOIA-2025-3933Form 483Closed
    2. Apr 15, 2025FDA-FOIA-2025-3420Form 483· 1 of 5 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 7 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000525253 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000525253. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.