Vortran Medical Technology 1, Inc
21 Golden Land Ct Ste 100, Sacramento, California, United States
FEI 1000307109
FDA has 3 inspections on file for Vortran Medical Technology 1, Inc in Sacramento, California (FEI 1000307109), from 2009 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection: 9 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2009–2017
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Statistical techniques - Lack of or inadequate procedures
Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.
Equipment suitability & capability
Certain inspection, measuring and test equipment is not suitable for its intended purposes and capable of producing valid results.
Show 5 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been documented in the device history record.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Info evaluated to determine if event was reportable
The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
Show all 10
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.250Statistical techniques1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000307109 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000307109. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.