Skip to content

Wello Inc. warning letter, November 8, 2023

FDA's Center for Devices and Radiological Health issued this warning letter to Wello Inc. of Addison, TX on November 8, 2023 and posted it on November 14, 2023. FDA lists its subject as CGMP/QSR/Medical Devices/Misbranded.

Source: FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Read the full letter on fda.gov. FDAgent is independent of FDA.

Issued
Nov 8, 2023
By CDRH
Posted by FDA
Nov 14, 2023
6 days after issue
Subject
CGMP/​QSR/​Medical Devices/​Misbranded
As FDA lists it
FDA reference
MARCS-CMS 664806
FDA's case number

Violations FDA states in this letter

This letter states no violation sentence that cites a 21 CFR section FDAgent can quote. The full letter is below.

The facility

FDA's compliance action record for this letter (case 664806) names one facility. Each facility's FDA record on FDAgent has its full inspection history.

Wello, Inc.Addison, Texas, United States · FEI 3011214569
Facility record →

Inspections

  • 2023-06-21OAI7 Form 483 citations
  • 2019-10-10VAI7 Form 483 citations

What FDA cited at the June 21, 2023 inspection

The letter

FDA's text of the letter as posted on fda.gov. FDAgent leaves out the address block, the greeting and the copy list, which name the people it was sent to.

During an inspection of your firm located in Addison, Texas, on June 15 through June 21, 2023, an investigator from the United States Food and Drug Administration (FDA) determined that your firm manufactures (or markets) the welloStationX.

Under section 201(h) of the Federal Food, Drug, and Cosmetic Act (the Act), 21 U.S.C. § 321(h), this product is a device because it is intended for use in the diagnosis of disease or other conditions or in the cure, mitigation, treatment, or prevention of disease, or to affect the structure or any function of the body.

Misbranding under Section 502(c) for Failure to Have Compliant Labeling

The inspection revealed that the welloStationX devices are misbranded within the meaning of section 502(c) of the Act, 21 U.S.C. § 352(c), because a word, statement, or other information required by or under authority of section 519 of the Act, 21 U.S.C. § 360i, to appear on the label or labeling of the devices was not prominently placed thereon with such conspicuousness (as compared with other words, statements, designs, or devices, in the labeling) and in such terms as to render it likely to be read and understood by the ordinary individual under customary conditions of purchase and use. In particular, 21 CFR 801.20(a) – which was promulgated under authority of section 519 of the Act, among other provisions – requires, with exceptions not relevant here, that the label and device package of every medical device bear a unique device identifier (UDI) that meets the requirements of 21 CFR Part 801, subpart B, and 21 CFR Part 830. The labels of the welloStationX devices do not bear such a UDI.

Specifically, the labels for welloStationX do not include a device identifier within the meaning of 21 CFR 801.3 or 830.3, there is no UDI presented in easily readable plain-text on the labels of the welloStationX (see 21 CFR 801.40(a)(1)), and there is no UDI presented in machine-readable form that uses automatic identification and data capture (AIDC) technology on the labels of the welloStationX (see 21 CFR 801.40(a)(2)). Misbranding under Section 502(t)(2) for Failure to Submit Required Information to GUDID

In addition, the welloStationX devices are misbranded within the meaning of section 502(t)(2) of the Act, 21 U.S.C. § 352(t)(2), in that there was a failure or refusal to furnish any material or information required by or under section 519 of the Act, 21 U.S.C. § 360i, respecting these devices. In particular, 21 CFR 830.300(a) and 830.320(b) – both of which were promulgated under section 519 of the Act, among other provisions – require that the labeler of a device submit electronically to FDA’s Global Unique Device Identification Database (GUDID) the information required by 21 CFR Part 830, subpart E, for each version or model required to bear a UDI. FDA has determined that your firm causes a label to be applied to a device with the intent that the device will be commercially distributed without any subsequent replacement or modification of the label. Wello, Inc. is therefore a “labeler” within the meaning of 21 CFR 830.3 and has not submitted to GUDID any information required by 21 CFR Part 830, subpart E, respecting these devices.

The failure or refusal to furnish any notification or other material or information required by or under section 519 of the Act, 21 U.S.C. § 360i, also constitutes a prohibited act under section 301(q)(1)(B) of the Act, 21 U.S.C. § 331(q)(1)(B).

Your firm should take prompt action to address any violations identified in this letter. Failure to promptly adequately address this matter may result in regulatory action being initiated by the FDA without further notice. These actions include, but are not limited to, seizure, injunction, and civil money penalties.

Other federal agencies may take your compliance with the FD&C Act and its implementing regulations into account when considering the award of federal contracts.

Please notify this office in writing within fifteen business days from the date you receive this letter of the specific steps your firm has taken to address the noted violations, as well as an explanation of how your firm plans to prevent these violations, or similar violations, from occurring again. Include documentation of the corrections and/or corrective actions (which must address systemic problems) your firm has taken. If your firm’s planned corrections and/or corrective actions will occur over time, please include a timetable for implementation of those activities. If corrections and/or corrective actions cannot be completed within fifteen business days, state the reason for the delay and the time within which these activities will be completed. Your firm’s response should be comprehensive and address any violations included in this Warning Letter. If you believe that your products are not in violation of the FD&C Act, include your reasoning and any supporting information for our consideration as part of your response.

Your firm’s response should be sent to: Jessica Mu, Director of Compliance Branch, at oradevices3firmresponse@fda.hhs.gov Refer to CMS # 664806 when replying. If you have any questions about the contents of this letter, please contact: Raymond W. Brullo, Compliance Officer at 949-608-2918 or raymond.brullo@fda.hhs.gov

Finally, you should know that this letter is not intended to be an all-inclusive list of the violations at your firm’s facility. It is your firm’s responsibility to ensure compliance with applicable laws and regulations administered by FDA.

Courtney H. Lias, Ph.D. Director OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Shari J. Shambaugh Program Division Director Office of Medical Device and Radiological Health Division 3 West

Ask FDAgent about this warning letter

FDAgent reads the letter, the facility's inspections and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The letter, its text, dates, issuing office and subject come from FDA's warning letter pages.
  • FDAgent leaves out the letter's address block, greeting and copy list, which name the people it was sent to, and quotes each numbered violation sentence that cites 21 CFR.
  • Facilities come from FDA's compliance action record for the letter's case number, or the one FEI the letter names; inspections and citations come from FDA's Data Dashboard.
  • FDAgent is independent of FDA.