Stryker Instruments Form 483, May 2009
FDA issued this Form 483 to Stryker Instruments in Portage, Michigan (FEI 1811755) at the end of a device inspection that ended May 14, 2009. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- May 14, 2009
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Jan 17, 2024
- 5,361 days after the inspection
FDA's 10 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Training
Employees have not been adequately trained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
General
Complaint handling procedures for receiving complaints have not been implemented.
Sampling methods - Lack of or inadequate procedures
Procedures were not established for reviewing sampling methods for adequacy for their intended use.
Performed by qualified individual
A validated process was not performed by a qualified individual.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.
Design validation - simulated testing
The design was not validated using production units under actual or simulated use conditions.
Specific non-conforming product procedures
Procedures for addressing the identification, evaluation, and investigation of nonconforming product were not defined.
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 1811755, newest first.
- Form 483, July 2015FDA record date Jul 11, 2015 · Posted Jan 17, 2024
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
- FDAgent is independent of FDA.