Repro-Med Systems, Inc. Form 483, December 2016

FDA issued this Form 483 to Repro-Med Systems, Inc. in Chester, New York (FEI 1318360) at the end of a device inspection that ended December 16, 2016. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Dec 16, 2016
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Jan 17, 2024
2,588 days after the inspection
Facility
FEI 1318360
Chester, New York

FDA's 8 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Product rework documentation, DHR {see also 820.184}

    Rework and reevaluation activities have not been documented in the device history record.

  2. Report of risk to health

    A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

  3. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  4. Design verification - output does not meet input requirement

    Design verification does not confirm that design output meets design input requirements.

  5. Design changes - Lack of or Inadequate Procedures

    Procedures for design change have not been adequately established.

  6. Acceptable supplier records, inadequate records

    Records of acceptable suppliers have not been adequately established.

  7. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  8. Nonconforming product, Lack of or inadequate procedures

    Procedures have not been adequately established to control product that does not conform to specified requirements.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.