Integra LifeSciences Corp Form 483, August 2011
FDA issued this Form 483 to Integra LifeSciences Corp in Plainsboro, New Jersey (FEI 1121308) at the end of a device inspection that ended August 25, 2011. FDA classified the inspection Official Action Indicated (OAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Aug 25, 2011
- Devices · Compliance: Devices
- FDA's outcome
- OAI
- Official Action Indicated
- Posted by FDA
- Jan 17, 2024
- 4,528 days after the inspection
FDA's 8 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Buildings
Buildings are not of suitable design to perform necessary operations.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
After this inspection
FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.
- Warning LetterDec 21, 2011CDRH · 118 days after this inspection ended
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.