American Contract Systems, Inc Form 483, August 2019

FDA issued this Form 483 to American Contract Systems, Inc in Temple Terrace, Florida (FEI 3003309947) at the end of a device inspection that ended August 16, 2019. FDA classified the inspection Official Action Indicated (OAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Nov 6, 2026. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Aug 16, 2019
Devices · Compliance: Devices
FDA's outcome
OAI
Official Action Indicated
Posted by FDA
Nov 6, 2026
2,639 days after the inspection
Facility
FEI 3003309947
Temple Terrace, Florida

FDA's 8 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  2. Software validation for automated processes

    Software used as part of production has not been adequately validated for its intended use according to an established protocol.

  3. Lack of or inadequate complaint procedures

    Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. Design validation- Lack of or inadequate procedures

    Procedures for design validation have not been established.

  5. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  6. Design control - no procedures

    Procedures for design control have not been established.

  7. Purchasing controls, Lack of or inadequate procedures

    Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  8. Design validation - Risk analysis

    Results of the design risk analysis were not documented.

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
  • FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.
  • FDAgent is independent of FDA.