FDA has 2 inspections on file for Baylab USA, LLC in Dallas, Texas (FEI 3017517827), from 2024 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 6 citations from this inspection. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.

FDA issued a warning letter on (CDRH · OHT4: Office of Surgical and Infection Control Devices, Office of Product Evaluation and Quality). No court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2024–2025
2 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Apr 1, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. GUDID Information required for unique device identification.

      The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device, the Global Medical Device Nomenclature (GMDN) term or code for the device and the total number of individual devices contained in the device package w

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

    Show 1 more citation
    1. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Label to bear a unique device identifier

      The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.

    4. GUDID Information required for unique device identification.

      The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device and the Global Medical Device Nomenclature (GMDN) term or code for the device was not provided to the FDA via the Global Unique Device Identification D

    5. Design control - no procedures

      Procedures for design control have not been established.

    Show 4 more citations
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained and implemented.

    2. Quality Audits - defined intervals

      Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Apr 1, 2026. Each one asked for a Form 483.

  • Inspection ending May 20, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jan 8, 2026. FDAgent holds no copy that FDA posted.

    1. Jun 25, 2025FDA-FOIA-2025-8474Form 483· 1 of 5 sitesRecords released
  • Inspection ending Jan 31, 2024OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 10, 2026. FDAgent holds no copy that FDA posted.

    1. Apr 1, 2026FDA-FOIA-2026-3167Form 483Records released
    2. Aug 8, 2024FDA-FOIA-2024-7108Form 483· 1 of 5 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 9

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3017517827 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3017517827. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.