Baylab USA, LLC
2230 Lbj Fwy Ste 100, Dallas, Texas, United States
FEI 3017517827
FDA has 2 inspections on file for Baylab USA, LLC in Dallas, Texas (FEI 3017517827), from 2024 to 2025. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 6 citations from this inspection. They are listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH · OHT4: Office of Surgical and Infection Control Devices, Office of Product Evaluation and Quality). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2024–2025
- 2 OAI · 0 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Apr 1, 2026
- Warning letters
- · CDRH · OHT4: Office of Surgical and Infection Control Devices, Office of Product Evaluation and Quality ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
GUDID Information required for unique device identification.
The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device, the Global Medical Device Nomenclature (GMDN) term or code for the device and the total number of individual devices contained in the device package w
Design control - no procedures
Procedures for design control have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
Show 1 more citation
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
GUDID Information required for unique device identification.
The correct unique device identifier (UDI) information regarding device identifier portion of the UDI assigned to the version or model, proprietary, trade, or brand name of the device as it appears on the label, size of the particular version or model given the device is available in more than one size, the type of production identifiers that appears on the label of the device, the FDA premarket submission number of a cleared or approved device and the Global Medical Device Nomenclature (GMDN) term or code for the device was not provided to the FDA via the Global Unique Device Identification D
Design control - no procedures
Procedures for design control have not been established.
Show 4 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained and implemented.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Apr 1, 2026. Each one asked for a Form 483.
Inspection ending May 20, 2025OAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jan 8, 2026. FDAgent holds no copy that FDA posted.
- Jun 25, 2025FDA-FOIA-2025-8474Form 483· 1 of 5 sitesRecords released
Inspection ending Jan 31, 2024OAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 10, 2026. FDAgent holds no copy that FDA posted.
- Apr 1, 2026FDA-FOIA-2026-3167Form 483Records released
- Aug 8, 2024FDA-FOIA-2024-7108Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 801.20Label to bear a unique device identifier2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 2
- 21 CFR 820.22Quality audit2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 830.310Information required for unique device identification2 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.75Process validation1 of 2
Show all 9
- 21 CFR 820.100Corrective and preventive action1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · OHT4: Office of Surgical and Infection Control Devices, Office of Product Evaluation and Quality · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3017517827 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3017517827. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.