WISSNER-BOSSERHOFF GMBH
Hauptstr. 4-6, Wickede (Ruhr), Germany
FEI 3015979371
FDA has one inspection on file for WISSNER-BOSSERHOFF GMBH in Wickede (Ruhr), Germany (FEI 3015979371). It ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 10 citations from this inspection: 9 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 1, in 2024
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jan 6, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.
Design output - documentation
Design output was not documented before release.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of the quality system have not been documented.
Show 5 more citations
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups and reviews by a designated person was not included in the record.
Production processes
Production processes were not developed to ensure that a device conforms to its specifications.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jan 6, 2025. It asked for a Form 483.
Inspection ending Apr 25, 2024OAI
1 request. In the inspection history above.
- Jan 6, 2025FDA-FOIA-2025-179Form 483Open
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 1
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3015979371 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3015979371. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.