FDA has 2 inspections on file for Fischer ANalysen Instrumente GmbH in Leipzig, Germany (FEI 3010837197), from 2022 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Nov 4, 2022. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
2, 2022–2024
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.

    3. Service reports

      Service reports do not include the required information.

    4. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 2 more citations
    1. Software validation for automated processes

      Software used as part of the quality system has not been validated for its intended use according to an established protocol.

    2. Equipment control activity documentation

      Equipment inspections, checks and maintenance activities have not been documented.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010837197 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010837197. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.