Skip to content

Advanced MicroVas II, LLC

8113 S Harvard Ave, Tulsa, Oklahoma, United States

FEI 3010281587

FDA has 2 inspections on file for Advanced MicroVas II, LLC in Tulsa, Oklahoma (FEI 3010281587), from 2013 to 2015. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2013–2015
2 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. DMR - not or inadequately maintained

      A device master record has not been maintained.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    5. Investigation of device failures

      Complaints involving the possible failure of a device and packaging to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    Show 6 more citations
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

    3. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    4. Personnel

      Personnel do not have the necessary education, background, training, and experience to perform their jobs.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    6. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    Show 6 more citations
    1. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    3. Design control - no procedures

      Procedures for design control have not been established.

    4. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    6. DMR - not or inadequately maintained

      A device master record has not been maintained.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010281587 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010281587. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.