Heat In A Click LLC
1975 Tigertail Blvd, Dania, Florida, United States
FEI 3008929787
FDA has one inspection on file for Heat In A Click LLC in Dania, Florida (FEI 3008929787). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Sep 29, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and consultants have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been established.
Show 5 more citations
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of procedures, or not maintained
Document control procedures have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 29, 2025. Each one asked for a Form 483.
Inspection ending Jan 24, 2025VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 15, 2025. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11404Form 483· 1 of 4 sitesRecords released
- Mar 7, 2025FDA-FOIA-2025-2143Form 483· 1 of 2 sitesClosed
- Mar 3, 2025FDA-FOIA-2025-1916Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 801.20Label to bear a unique device identifier1 of 1
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.150Storage1 of 1
Show all 10
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008929787 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008929787. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.