Psoria Shield, Inc.
409 Mandeville St, Utica, New York, United States
FEI 3008872264
FDA has 4 inspections on file for Psoria Shield, Inc. in Utica, New York (FEI 3008872264), from 2012 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.198(a) and 21 CFR 820.72(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2012–2025
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Dec 22, 2023
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and maintained have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Show 1 more citation
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Certification label or tag
A certification label or tag is not in the English language, permanently affixed or inscribed, and readily accessible to view when the product is fully assembled for use.
Rpts of safety & effectiveness, adulteration or misbranding
The manufacturer failed to provide other information bearing on the classification of a device and indicating whether it is adulterated or misbranded under the act.
Results of tests
You have not maintained records of the results of tests for electronic product radiation safety, including the control of unnecessary, secondary or leakage electronic product radiation.
Reports and information documentation
The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Show 1 more citation
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Dec 22, 2023. Each one asked for a Form 483.
Inspection ending Sep 14, 2023NAI
2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Dec 22, 2023FDA-FOIA-2023-11438Form 483Closed
- Oct 16, 2023FDA-FOIA-2023-9037Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 1002.10Product reports1 of 3
Show all 11
- 21 CFR 1002.30Records to be maintained by manufacturers1 of 3
- 21 CFR 1010.2Certification1 of 3
- 21 CFR 860.7Determination of safety and effectiveness1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008872264 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008872264. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.