Skip to content

American Contract Systems Inc

85 Shaffer Park Dr, Tiffin, Ohio, United States

FEI 3008492337

FDA has 5 inspections on file for American Contract Systems Inc in Tiffin, Ohio (FEI 3008492337), from 2011 to 2025. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the four FOIA requests for this site has asked FDA for this inspection's 483.

FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.80(b) and 21 CFR 820.75(a).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
5, 2011–2025
2 OAI · 1 VAI · 2 NAI
Inspections with 483 observations shown
3 of 5
Form 483s posted by FDA
↗
Requested under FOIA
4 requests, latest Jun 30, 2026
Warning letters
· CDRH ↗
2 on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sampling methods - Lack of or inadequate procedures

      Procedures to ensure sampling methods are adequate for their intended use have not been adequately established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 4 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Documentation

      Process validation activities and results have not been adequately documented.

    3. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    4. Design history file

      The design history file was not established and does not demonstrate that the design was developed following the approved design plan.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Documentation

      Acceptance activities were not adequately documented.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on Jun 30, 2026. Each one asked for a Form 483.

  • Inspection ending Jun 2, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 2, 2025. FDAgent holds no copy that FDA posted.

    1. Jul 31, 2025FDA-FOIA-2025-9272Form 483· 1 of 5 sitesRecords released
  • Inspection ending Aug 14, 2024NAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. Jun 30, 2026FDA-FOIA-2026-6208Form 483· 1 of 5 sitesRecords released
    2. Nov 8, 2024FDA-FOIA-2024-9926Form 483· 1 of 2 sitesClosed
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. Nov 20, 2024FDA-FOIA-2024-10246Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008492337 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008492337. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.