FDA has 5 inspections on file for Carefusion 2200 Inc in Vernon Hills, Illinois (FEI 3008002452), from 2011 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection: 1 under 21 CFR Part 806, the device correction and removal reporting rules, and 1 under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Nov 5, 2013. FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.100(a) and 21 CFR 803.50(a)(2).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2011–2022
1 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Oct 3, 2024
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Event description

      The events that led to reporting and/or conducting the correction or removal actions were not described.

  2. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been adequately established.

    2. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers were not evaluated and selected based on their ability to meet specified requirements.

    5. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    Show 2 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Oct 3, 2024. It asked for a Form 483.

  • Inspection ending Oct 19, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 3, 2024FDA-FOIA-2024-8679Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3008002452, one row per refused product line.

1 refusal on 2018-08-23 · charged NOT LISTED on 1, REGISTERED on 1 · no FDA sample analysis

FDA firm profile
  • 2018-08-23

    CATHETER, PERCUTANEOUS74DQY

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-7266025-9/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3008002452 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3008002452. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.