Skip to content

CardioQuip, LLC

8422 Calibration Ct, College Station, Texas, United States

FEI 3007899424

FDA has 3 inspections on file for CardioQuip, LLC in College Station, Texas (FEI 3007899424), from 2015 to 2021. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 803.17, 21 CFR 820.198(a) and 21 CFR 820.30(i).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
3, 2015–2021
1 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Feb 26, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate and is incomplete.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    5. Records of MDR Investigation

      Investigation records of MDR reportable complaints do not include required information.

    Show 6 more citations
    1. Design output - documentation

      Design output was not adequately documented before release.

    2. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    6. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    5. Acceptable supplier records, inadequate records

      Records of acceptable contractors have not been adequately established.

    Show 1 more citation
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and implemented.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Feb 26, 2026. It asked for a Form 483.

  • Inspection ending Sep 17, 2021OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 10, 2026. FDAgent holds no copy that FDA posted.

    1. Feb 26, 2026FDA-FOIA-2026-1994Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007899424 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007899424. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.