Skip to content

CellEra LLC

1045 Reed Dr Ste C, Monroe, Ohio, United States

FEI 3007594734

FDA has 3 inspections on file for CellEra LLC in Monroe, Ohio (FEI 3007594734), from 2019 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

The most recent OAI inspection ended on Jul 8, 2021. FDA issued a warning letter on (CDRH · Office of Medical Device and Radiological Health Division 1 - East). No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2019–2022
1 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Oct 4, 2023
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    2. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Design control - no procedures

      Procedures for design control have not been established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    Show 3 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    3. Lack of procedures, or not maintained

      Document control procedures have not been established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    5. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    Show 2 more citations
    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    2. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Oct 4, 2023. Each one asked for a Form 483.

  • Inspection ending Jul 8, 2021OAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 4, 2023FDA-FOIA-2023-8608Form 483· 1 of 8 sitesClosed
    2. Jun 1, 2022FDA-FOIA-2022-4022Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007594734 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007594734. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.