Neo Medical, Inc.
1375 Greg St Ste 108, Sparks, Nevada, United States
FEI 3007207906
FDA has 6 inspections on file for Neo Medical, Inc. in Sparks, Nevada (FEI 3007207906), from 2009 to 2014. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Mar 26, 2010. No warning letter, court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.70(g)(2).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2009–2014
- 1 OAI · 1 VAI · 4 NAI
- Inspections with 483 observations shown
- 2 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Use and removal, Lack of or inadequate procedures
Procedures for the use and removal of manufacturing material have not been adequately established.
Periodic equipment inspections
Periodic inspections of equipment conducted to ensure adherence to applicable maintenance schedules were not documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Contamination control, Lack of or inadequate procedures
Procedures have not been defined to prevent contamination of equipment or product by certain substances.
Periodic equipment inspection lack of or inadequate procedu
Periodic inspections of equipment conducted in accordance with established procedures to ensure adherence to applicable maintenance schedules were not documented.
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals.
Lack of or inadequate In-process acceptance procedures
Acceptance procedures to ensure that specified requirements for in-process product are met were not implemented.
Show 3 more citations
Calibration procedures - content
Calibration procedures do not include specific directions and limits for accuracy and precision.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls2 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.20Management responsibility1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.198Complaint files1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007207906 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007207906. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.