Eizo GmbH
Carl-Benz Str. 3, 76761, Rulzheim, Germany
FEI 3006615345
FDA has 4 inspections on file for Eizo GmbH in Rulzheim, Germany (FEI 3006615345), from 2012 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
2 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.30(g) and 21 CFR 820.80(d).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2012–2023
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Show 1 more citation
Service reports
Service reports do not include the required information.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Design validation - documentation
The results of design validation, including identification of the design, method(s), and the date, were not documented in the design history file.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design plans- updated
Design plans were not updated as design and development evolves.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.120Device labeling1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 820.200Servicing1 of 3
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3006615345, one row per refused product line.
2 refusals on 2021-07-16 · charged FRNMFGREG on 2, REGISTERED on 2 · no FDA sample analysis
FDA firm profile- 2021-07-16
DISPLAY, CATHODE-RAY TUBE, MEDICAL74DXJ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-6157674-2/11/1
- 2021-07-16
DISPLAY, CATHODE-RAY TUBE, MEDICAL74DXJ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-6154777-6/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006615345 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006615345. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.