OrthoPediatrics Corp
2850 Frontier Dr, Warsaw, Indiana, United States
FEI 3006460162
FDA has 5 inspections on file for OrthoPediatrics Corp in Warsaw, Indiana (FEI 3006460162), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below. Three FOIA requests have asked FDA for this inspection's 483.
13 import refusals are filed under this FEI, the latest on . No warning letter or court action is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.198(a) and 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2023
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Warning letters
- None on file
- Import refusals
- 13 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not implemented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Aug 18, 2023VAI
3 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-8006Form 483Closed
- Oct 19, 2023FDA-FOIA-2023-9268Form 483Closed
- Oct 18, 2023FDA-FOIA-2023-9208Form 483Denied
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 6 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 806.10Reports of corrections and removals1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3006460162, one row per refused product line.
13 refusals on 2019-06-03 · charged FRNMFGREG on 13, REGISTERED on 13 · no FDA sample analysis
FDA firm profile- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/91/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/81/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/71/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/61/1
- 2019-06-03
SCREW, FIXATION, BONE87HWC
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/51/1
Show 8 more
- 2019-06-03
SCREW, FIXATION, BONE87HWC
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/41/1
- 2019-06-03
SCREW, FIXATION, BONE87HWC
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/31/1
- 2019-06-03
SCREW, FIXATION, BONE87HWC
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/21/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/131/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/121/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/111/1
- 2019-06-03
SCREW, FIXATION, BONE87HWC
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/11/1
- 2019-06-03
PLATE, FIXATION, BONE87HRS
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports336-7951757-8/101/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006460162 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006460162. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.