Contacare Ophthalmics & Diagnostics
310b Village Sim, Block No, Dabhasa, India
FEI 3006217304
FDA has 3 inspections on file for Contacare Ophthalmics & Diagnostics in Dabhasa, India (FEI 3006217304), from 2013 to 2018. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDER).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 11 citations from this inspection, all under 21 CFR Part 211, the manufacturing rules for finished drugs. They are listed below.
7 import refusals are filed under this FEI, the latest on . FDA’s Red Lists for Import Alerts 55-05 and 66-40 name this firm at this address. FDA printed no FEI on the entries. No court action is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2013–2018
- 2 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDER ↗
- Import refusals
- 7 refusals
- Import alerts
- 55-05, 66-40Red List
- Matched on FDA’s firm name and street address
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established, written and followed.
Deviations from laboratory control requirements
Deviations from written specifications and test procedures are not justified.
Environmental Monitoring System
Aseptic processing areas are deficient regarding the system for monitoring environmental conditions.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards and test procedures designed to assure that components, drug product containers, closures, in-process materials, labeling and drug products conform to appropriate standards of identity, strength, quality and purity.
Aseptic Processing Area
Separate or defined areas to prevent contamination or mix-ups are deficient regarding operations related to aseptic processing of drug products.
Show 6 more citations
SOPs not followed / documented
Written production and process control procedures are not documented at the time of performance.
Validation lacking for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile did not include adequate validation of the aseptic and sterilization process.
Complaint Handling Procedure
Procedures describing the handling of written and oral complaints related to drug products are not written or followed.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel. Electronic records are used, but they do not meet system access limitation, employee accountability/responsibility policy and systems documentation control requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.
Extent of discrepancy, failure investigations
Investigations of an unexplained discrepancy and a failure of a batch or any of its components to meet any of its specifications did not extend to other batches of the same drug product.
Cleaning System
Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the room and equipment to produce aseptic conditions.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Prevention of cross contamination, mix-ups
There is insufficient physical or spatial separation from operations and other drug products to prevent mix-ups and cross-contamination.
Complaint Investigation/Follow-Up Findings
Complaint records are deficient in that they do not include the findings of the investigation.
Written calibration / inspection records not kept
Records of the calibration checks of automatic, mechanical or electronic equipment, including computers or related systems are not maintained.
Procedure Deviations Recorded and Justified
Deviations from written production and process control procedures are not recorded.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training , Education , Experience overall
Employees engaged in the processing, packing, and holding of a drug product lack the training required to perform their assigned functions.
Valid stability test methods
The written stability program for drug products does not include specific test methods.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and followed.
Written procedures not established/followed
Written procedures are not established and followed for the cleaning and maintenance of equipment, including utensils, used in the manufacture, processing, packing or holding of a drug product.
Laboratory equipment calibration records
Laboratory records do not include complete records of the periodic calibration of laboratory instruments.
Show 14 more citations
Test methods
The specificity of test methods have not been established.
Prepared for each batch, include complete information
Batch production and control records do not include complete information relating to the production and control of each batch.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions.
Complete test data included in records
Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards.
GMP Training Frequency
GMP training is not conducted on a continuing basis and with sufficient frequency to assure that employees remain familiar with CGMP requirements applicable to them.
Scientifically sound laboratory controls
Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, standards, and test procedures designed to assure that components and drug products conform to appropriate standards of identity, strength, quality and purity.
Written sanitation procedures not followed
Written procedures for sanitation are not followed.
Reports of Analysis (Components)
Reports of analysis from component suppliers are accepted in lieu of testing each component for conformity with all appropriate written specifications, without establishing the reliability of the supplier's analyses through appropriate validation of the supplier's test results at appropriate intervals.
Deviations from laboratory control requirements
Deviations from written test procedures are not justified.
Complaint Investigation/Follow-Up Findings
Complaint records are deficient in that they do not include the findings of the investigation and follow-up.
Second person sign off
Laboratory records do not include the initials or signature of a second person showing that the original records have been reviewed for accuracy, completeness, and compliance with established standards.
Written in-process control procedures
Written procedures are not established that describe the examinations to be conducted on appropriate samples of in-process materials of each batch.
Following/documenting laboratory controls
Established laboratory control mechanisms are not followed and documented at the time of performance.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not fully followed.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 211.100Written procedures; deviations3 of 3
- 21 CFR 211.198Complaint files3 of 3
- 21 CFR 211.68Automatic, mechanical, and electronic equipment2 of 3
- 21 CFR 211.113Control of microbiological contamination2 of 3
- 21 CFR 211.160General requirements2 of 3
- 21 CFR 211.22Responsibilities of quality control unit1 of 3
- 21 CFR 211.25Personnel qualifications1 of 3
- 21 CFR 211.42Design and construction features1 of 3
Show all 18
- 21 CFR 211.56Sanitation1 of 3
- 21 CFR 211.67Equipment cleaning and maintenance1 of 3
- 21 CFR 211.84Testing and approval or rejection of components, drug product containers, and closures1 of 3
- 21 CFR 211.110Sampling and testing of in-process materials and drug products1 of 3
- 21 CFR 211.130Packaging and labeling operations1 of 3
- 21 CFR 211.165Testing and release for distribution1 of 3
- 21 CFR 211.166Stability testing1 of 3
- 21 CFR 211.188Batch production and control records1 of 3
- 21 CFR 211.192Production record review1 of 3
- 21 CFR 211.194Laboratory records1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3006217304, one row per refused product line.
7 refusals from 2017-06-23 to 2025-01-22 · charged DRUG GMPS on 3, NOT LISTED on 1 · FDA sample analysis on 1
FDA firm profile- 2025-01-22
AID, SURGICAL, VISCOELASTIC86LZP
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-5230813-9/11/1
- 2018-05-02
STRABISMUS DETECTION DEVICE86PMW
INVDEVICE 501(i), 801(a)(3); ADULTERATION
Division of Southeast Imports799-5507090-6/11/1
- 2018-01-19
FLUORESCEIN SODIUM (DIAGNOSTIC AID) (DRUGS)63XBD12
DRUG GMPS 501(a)(2)(B), 801(a)(3); ADULTERATION
Division of Northern Border ImportsUPS-6572371-1/11/1
- 2018-01-19
FLUORESCEIN SODIUM (DIAGNOSTIC AID) (DRUGS)63XAY12
DRUG GMPS 501(a)(2)(B), 801(a)(3); ADULTERATION
Division of Northern Border Imports799-3938049-6/11/1
- 2018-01-19
FLUORESCEIN SODIUM (DIAGNOSTIC AID) (DRUGS)63XAY12
DRUG GMPS 501(a)(2)(B), 801(a)(3); ADULTERATION
Division of Northern Border Imports799-3914132-8/11/1
Show 2 more
- 2018-01-09
FLUORESCEIN SODIUM (DIAGNOSTIC AID) (DRUGS)63XAY12
FILTH 501(a)(1)Adulteration
Division of Northern Border Imports799-3977242-9/11/1
- 2017-06-23
SODIUM CHLORIDE (REPLENISHER)65RAK19
STERILITY 501(a)(2)(A), 801(a)(3); ADULTERATION
FDA sample analysis
Division of Southwest ImportsDM4-0487577-3/11/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 66-40Red ListDetention Without Physical Examination of Drugs From Firms Which Have Not Met Drug GMPsImport alert pageAlert on fda.gov
Date published 2019-02-20
- FDA’s entry
- Contacare Ophthalmics & Diagnostics · 310b Village Sim, Block No, Dabhasa, Gujarat INDIA
- FDA’s inspection record for FEI 3006217304
- Contacare Ophthalmics & Diagnostics · 310b Village Sim, Block No, Dabhasa, India
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
10 products
- 55 - - - --Pharm Necess & Ctnr For Drug/Bio
- 56 - - - --Antibiotics (Human/Animal)
- 58 - - - --Human and Animal Therapeutic Biologic and Biosimilar Drugs
Show 7 more products
- 60 - - - --Human and Animal Drugs
- 61 - - - --Human and Animal Drugs
- 62 - - - --Human and Animal Drugs
- 63 - - - --Human and Animal Drugs
- 64 - - - --Human and Animal Drugs
- 65 - - - --Human and Animal Drugs
- 66 - - - --Human and Animal Drugs
- 55-05Red ListDetention Without Physical Examination Of Finished Dosage Form Drug Products, Active Pharmaceutical Ingredients, and Inactive Ingredients for Potentially Hazardous Microbiological ContaminationImport alert pageAlert on fda.gov
Date published 2017-08-23
- FDA’s entry
- Contacare Ophthalmics & Diagnostics · 310b Village Sim, Block No, Dabhasa, Gujarat INDIA
- FDA’s inspection record for FEI 3006217304
- Contacare Ophthalmics & Diagnostics · 310b Village Sim, Block No, Dabhasa, India
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
7 products
- 55 R P - 14Hydroxypropyl Methylcellulose (Pharmaceutic Necessity - Suspending Agent), Pharma Necess
- 55 R P - 55Sodium Chloride (Pharmaceutic Necessity - Tonicity Agent), Pharma Necess
- 63 X C P 69Iopydol (Diagnostic Aid) (Drugs), Human - Rx/Single Ingredient, Small Volume Parenteral <100ml
Show 4 more products
- 63 X - - 12Fluorescein Sodium (Diagnostic Aid) (Drugs)
- 63 X - - 99Diagostic Aid (Drugs) N.E.C.
- 65 R A K 19Sodium Chloride (Replenisher), Human - Non/Rx Single Ingredient, Sterile Liquid
- 65 R D K 12Potassic Saline (Injection) (Replenisher), Human - Rx/Combination Ingredient, Sterile Liquid
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006217304 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006217304. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.