Medart A /S

Industriholmen 15a, Hvidovre, Denmark

FEI 3005616124

FDA has 3 inspections on file for Medart A /S in Hvidovre, Denmark (FEI 3005616124), from 2012 to 2015. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Jun 1, 2012. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.30(e).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2012–2015
1 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
2 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Aperture label - laser radiation

      The warning statement "AVOID EXPOSURE---Laser radiation is emitted from this aperture" was not affixed in close proximity to each aperture through which is emitted accessible laser or collateral radiation in excess of the accessible emission limits of Class I and table VI of Section 1040.10(d).

    2. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    3. Failure to submit, distinct marking

      You did not submit a required product report prior to the introduction of the product into commerce.

    4. Design review procedures - personnel

      Procedures were not established for ensuring that participants at each design review include an individual who does not have direct responsibility for the design stage being reviewed.

    5. Supplier notification of changes

      There is no clear agreement from the suppliers, contractors, and consultants that they will notify you of changes in the product or service.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.

    2. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

    3. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    4. Design review - Lack of or inadequate procedures

      Procedures for design review have not been adequately established.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    Show 7 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Statistical techniques - Lack of or inadequate procedures

      Procedures for identifying valid statistical techniques required for establishing, controlling, and verifying the acceptability of process capability and product characteristics have not been adequately established.

    4. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    5. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    6. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    7. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.

  • 21 CFR 820.30Design controls2 of 2
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 2
  • 21 CFR 820.20Management responsibility1 of 2
  • 21 CFR 820.25Personnel1 of 2
  • 21 CFR 820.50Purchasing controls1 of 2
  • 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
  • 21 CFR 820.100Corrective and preventive action1 of 2
Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005616124 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005616124. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.