Skip to content

Nouvag Ag

St. Gallerstrasse 23-25, Goldach, Switzerland

FEI 3005601686

FDA has 2 inspections on file for Nouvag Ag in Goldach, Switzerland (FEI 3005601686), from 2016 to 2018. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

8 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.50 and 21 CFR 820.75(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2016–2018
1 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
2 of 2
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
8 refusals
Import alerts
89-08Red List
Matched on FDA’s firm name and street address

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Records of complaint investigation

      Records of complaint investigations do not include required information.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    5. Identification procedures, Lack of or inadequate procedures

      Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been established.

    Show 1 more citation
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3005601686, one row per refused product line.

8 refusals from 2019-04-16 to 2026-08-23 · charged NO 510(K) on 5, NOT LISTED on 5 · no FDA sample analysis

FDA firm profile
  • 2026-08-23

    PUMP, INFUSION80FRN

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports1FX-2582065-5/10002/1

  • 2024-12-10

    SYSTEM, SUCTION, LIPOPLASTY79MUU

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of West Coast ImportsE7T-0902241-6/11/1

  • 2022-05-11

    MOTOR, SURGICAL INSTRUMENT, AC-POWERED79GEY

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast ImportsSCS-1335082-5/11/1

  • 2020-07-13

    TUBING, NONINVASIVE80GAZ

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southwest ImportsDM0-0992203-8/21/1

  • 2019-07-25

    CANNULA, NASAL, OXYGEN73CAT

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-2368678-4/11/3

Show 3 more
  • 2019-07-25

    CANNULA, NASAL, OXYGEN73CAT

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-2368678-4/11/2

  • 2019-07-25

    APPARATUS, AUTOTRANSFUSION73CAC

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-2368678-4/11/1

  • 2019-04-16

    SYSTEM, SUCTION, LIPOPLASTY79MUU

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast ImportsAEK-5827611-1/11/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov

    Date published 2019-08-14

    FDA’s entry
    Nouvag Ag · St. Gallerstrasse 23-25, Goldach, St. Gallen SWITZERLAND
    FDA’s inspection record for FEI 3005601686
    Nouvag Ag · St. Gallerstrasse 23-25, Goldach, Switzerland

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    3 products

    • 79 J - - CXApparatus, Suction, Ward Use, Portable, Ac-Powered
    • 79 M - - UUSystem, Suction, Lipoplasty
    • 80 F - - RNPump, Infusion

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3005601686 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3005601686. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.