Radlink Inc
2101 E El Segundo Blvd, ste 104, El Segundo, California, United States
FEI 3005560996
FDA has 4 inspections on file for Radlink Inc in El Segundo, California (FEI 3005560996), from 2009 to 2018. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2009–2018
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 1 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not documented and implemented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure that equipment is routinely calibrated were not implemented.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not implemented.
Lack of procedures, or not maintained
Document control procedures were not implemented.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Show 5 more citations
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been completed.
Design control - no procedures
Procedures to control the design process of the device were not implemented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not defined and implemented.
Resources
Adequate resources have not been provided for performing management activities, assessment activities, and audits.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 4 inspections.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.25Personnel1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
Show all 9
- 21 CFR 820.198Complaint files1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3005560996 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3005560996. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.