FEI 3004753838

FDA has 11 inspections on file for Dexcom, Inc. in San Diego, California (FEI 3004753838), from 2008 to 2024. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Seven FOIA requests have asked FDA for this inspection's records, and one more for the 483 from a later inspection, ending Mar 4, 2025, that FDA's Data Dashboard doesn't list yet.

6 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.30(g), 21 CFR 820.198(a) and 21 CFR 803.17.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
11, 2008–2024
3 OAI · 3 VAI · 5 NAI
Inspections with 483 observations shown
6 of 11
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
8 requests, latest Nov 21, 2025
Warning letters
· CDRH ↗
3 on file
Import refusals
6 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design input - Lack of or inadequate procedures

      Procedures for design input have not been adequately established.

    2. Design output - Lack of or inadequate procedures

      Procedures for design output have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate and is incomplete.

    Show 2 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Sponsors' general responsibilities

      For an investigational study, proper monitoring was not ensured.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

  6. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  7. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed and maintained.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    5. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    Show 1 more citation
    1. Information used for annual reports

      The written MDR procedure does not include documentation and recordkeeping requirements for any information that was evaluated for the purpose of preparing the submission of semiannual reports.

  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  9. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  10. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  11. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - Lack of or inadequate procedures

      Design verification did not confirm that the design output meets the design input requirements.

    2. Environmental control system periodic inspections

      Environmental control systems have not been inspected periodically to verify that the system, including necessary equipment, is adequate and functioning properly.

    3. Design validation - software validation not performed

      Validation of device software is incomplete.

    4. Environmental control Lack of or inadequate procedures

      Appropriate procedures have not been followed for controlling environmental conditions.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Eight requests to FDA under the Freedom of Information Act named this site, most recently on Nov 21, 2025.

  • Inspection ending Mar 4, 2025

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Mar 20, 2025FDA-FOIA-2025-2573Form 483Closed
  • Inspection ending Nov 7, 2024OAI

    7 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 21, 2025. FDAgent holds no copy that FDA posted.

    1. Nov 21, 2025FDA-FOIA-2025-13222Form 483· 1 of 2 sitesOpen
    2. Aug 28, 2025FDA-FOIA-2025-10338Form 483Records released
    3. Aug 1, 2025FDA-FOIA-2025-9337Inspection records· 1 of 2 sitesRecords released
    4. Apr 30, 2025FDA-FOIA-2025-3887Form 483· 1 of 4 sitesClosed
    5. Mar 24, 2025FDA-FOIA-2025-2700Form 483Closed
    6. Mar 13, 2025FDA-FOIA-2025-2353EIRWithdrawn
    7. Mar 13, 2025FDA-FOIA-2025-2319Form 483Released in part

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 11 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004753838, one row per refused product line.

6 refusals from 2023-11-28 to 2026-10-08 · charged NOT LISTED on 4, NO 510(K) on 4 · no FDA sample analysis

FDA firm profile
  • 2026-10-08

    GLUCOSE, NONINVASIVE TECHNOLOGY75NCT

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports1FX-5677680-0/10001/1

  • 2025-06-05

    INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast ImportsSCS-0763403-8/12/1

  • 2025-05-27

    INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsSCS-0647837-9/12/1

  • 2024-10-08

    TAPE AND BANDAGE, ADHESIVE79KGX

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports799-3232106-1/11/1

  • 2023-12-13

    INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast ImportsSCS-1161893-5/11/1

Show 1 more
  • 2023-11-28

    INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Northern Border ImportsAEK-4311124-2/11/1

Other FEIs recorded as Dexcom, Inc.

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

  • Dexcom, Inc.FEI 3013435121Mesa, Arizona, United States

    Last inspected 2018-03-22NAI

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004753838 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004753838. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.