Dexcom, Inc.
6340 Sequence Dr, San Diego, California, United States
FEI 3004753838
FDA has 11 inspections on file for Dexcom, Inc. in San Diego, California (FEI 3004753838), from 2008 to 2024. The latest ended on and was classified OAI, official action indicated. FDA has issued 3 warning letters under this FEI, the latest on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Seven FOIA requests have asked FDA for this inspection's records, and one more for the 483 from a later inspection, ending Mar 4, 2025, that FDA's Data Dashboard doesn't list yet.
6 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.30(g), 21 CFR 820.198(a) and 21 CFR 803.17.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 11, 2008–2024
- 3 OAI · 3 VAI · 5 NAI
- Inspections with 483 observations shown
- 6 of 11
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 8 requests, latest Nov 21, 2025
- Import refusals
- 6 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
11 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Design output - Lack of or inadequate procedures
Procedures for design output have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Show 2 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Bioresearch Monitoring1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sponsors' general responsibilities
For an investigational study, proper monitoring was not ensured.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Project Evaluation: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and maintained.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Show 1 more citation
Information used for annual reports
The written MDR procedure does not include documentation and recordkeeping requirements for any information that was evaluated for the purpose of preparing the submission of semiannual reports.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Environmental control system periodic inspections
Environmental control systems have not been inspected periodically to verify that the system, including necessary equipment, is adequate and functioning properly.
Design validation - software validation not performed
Validation of device software is incomplete.
Environmental control Lack of or inadequate procedures
Appropriate procedures have not been followed for controlling environmental conditions.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Eight requests to FDA under the Freedom of Information Act named this site, most recently on Nov 21, 2025.
Inspection ending Mar 4, 2025
1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.
- Mar 20, 2025FDA-FOIA-2025-2573Form 483Closed
Inspection ending Nov 7, 2024OAI
7 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 21, 2025. FDAgent holds no copy that FDA posted.
- Nov 21, 2025FDA-FOIA-2025-13222Form 483· 1 of 2 sitesOpen
- Aug 28, 2025FDA-FOIA-2025-10338Form 483Records released
- Aug 1, 2025FDA-FOIA-2025-9337Inspection records· 1 of 2 sitesRecords released
- Apr 30, 2025FDA-FOIA-2025-3887Form 483· 1 of 4 sitesClosed
- Mar 24, 2025FDA-FOIA-2025-2700Form 483Closed
- Mar 13, 2025FDA-FOIA-2025-2353EIRWithdrawn
- Mar 13, 2025FDA-FOIA-2025-2319Form 483Released in part
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 11 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 6
- 21 CFR 820.30Design controls2 of 6
- 21 CFR 820.198Complaint files2 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 6
- 21 CFR 812.40General responsibilities of sponsors1 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.75Process validation1 of 6
- 21 CFR 820.100Corrective and preventive action1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3004753838, one row per refused product line.
6 refusals from 2023-11-28 to 2026-10-08 · charged NOT LISTED on 4, NO 510(K) on 4 · no FDA sample analysis
FDA firm profile- 2026-10-08
GLUCOSE, NONINVASIVE TECHNOLOGY75NCT
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports1FX-5677680-0/10001/1
- 2025-06-05
INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-0763403-8/12/1
- 2025-05-27
INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsSCS-0647837-9/12/1
- 2024-10-08
TAPE AND BANDAGE, ADHESIVE79KGX
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast Imports799-3232106-1/11/1
- 2023-12-13
INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsSCS-1161893-5/11/1
Show 1 more
- 2023-11-28
INTEGRATED CONTINUOUS GLUCOSE MONITORING SYSTEM, FACTORY CALIBRATED75QBJ
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
Division of Northern Border ImportsAEK-4311124-2/11/1
Other FEIs recorded as Dexcom, Inc.
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Dexcom, Inc.FEI 3013435121Mesa, Arizona, United States
Last inspected 2018-03-22NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3004753838 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3004753838. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.