Skip to content

Infinite Biomedical Technologies

8 Market Pl Ste 500, Baltimore, Maryland, United States

FEI 3004426795

FDA has 3 inspections on file for Infinite Biomedical Technologies in Baltimore, Maryland (FEI 3004426795), from 2019 to 2026. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection. They are listed below. Three of the four FOIA requests for this site have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2019–2026
0 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
4 requests, latest FY2023
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 7.3.3

      Design and development inputs do not include all requirements

      Design and development inputs related to product requirements do not adequately include functional, performance, usability, or safety requirements, according to the intended use, applicable output(s) of risk management and information derived from previous similar designs.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action not taken

      Corrective action was not adequately taken to eliminate the cause of nonconformities in order to prevent recurrence.

    3. ISO 13485:2016 Clause 6.1(b)

      Resources for meeting regulatory and customer requirements not provided

      Resources needed to meet applicable regulatory and customer requirements have not been adequately determined and provided.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers, contractors and consultants have not been adequately established.

    3. Software validation for automated processes

      Software used as part of the quality system has not been adequately validated for its intended use according to an established protocol.

    4. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Four requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.

  • Inspection ending Jul 30, 2026VAI

    3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 24, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 24, 2026FDA-FOIA-2026-9251Form 483· 1 of 5 sitesOpen
    2. Sep 10, 2026FDA-FOIA-2026-8787Form 483Records released
    3. Sep 1, 2026FDA-FOIA-2026-8485Form 483· 1 of 5 sitesRecords released
  • Inspection ending Jun 10, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. FY2023FDA-FOIA-2022-7437Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004426795 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004426795. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.