Skip to content

Stembanc Inc

100 7th Ave Ste 200, Chardon, Ohio, United States

FEI 3004195871

FDA has 6 inspections on file for Stembanc Inc in Chardon, Ohio (FEI 3004195871), from 2009 to 2024. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Six regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2009–2024
0 OAI · 4 VAI · 2 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upBiologics · Human Cellular, Tissue, and Gene TherapiesNone
  2. NAINo objectionable conditions that need follow-upBiologics · Vaccines and Allergenic ProductsNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Validation & approval--established procedures

      Processes with results which could not be fully verified by inspection and tests, were not validated and approved according to established procedures.

    2. Quality audits performed periodically

      Periodic quality audits of activities related to core CGTP requirements have not been performed.

    3. Procedures and release criteria

      Procedures including release criteria for activities relating to the distribution of HCT/Ps were not revised.

    4. Causing contamination, increasing risks

      HCT/Ps were not processed in a way that does not cause contamination or cross contamination during processing and that does not increase the risk of introduction, transmission, or spread of communicable disease.

    5. Deviations--evaluation, cause, corrective action

      Investigation of deviations related to core CGTP requirements did not include efforts to determine the cause of the deviation and corrective action(s) to address the deviation and prevent recurrence.

    Show 1 more citation
    1. Birth mother sample not tested

      A sample from the birth mother of a donor one month of age or younger was not used for testing of communicable disease agents.

  4. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Autologous use only--labeling

      HCT/Ps for autologous use and for which the donor eligibility determination was not performed were not labeled for autologous use only.

    2. Eligibility not required--warning labels

      HCT/Ps for which the donor eligibility determination was not performed were not prominently labeled with the appropriate warning statements.

    3. Listing and interpretation of CD tests performed

      The summary of records for HCT/Ps did not contain of all communicable disease tests performed.

    4. Birth mother sample not tested

      A sample from the birth mother of a donor one month of age or younger was not used for testing of communicable disease agents.

    5. All labeling requirements met

      Labeling procedures did not ensure that HCT/Ps are labeled in accordance with all labeling requirements.

  5. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Manufacturer instructions not followed

      Testing for communicable disease agents was not performed in accordance with the manufacturer's instructions.

    2. Improper release prevention--storage, I.D.

      HCT/Ps from ineligible donors are not stored and identified in a manner to prevent improper release.

    3. Training of personnel

      The quality program has not ensured the proper training and education of personnel involved in core GTP activities.

    4. Risk factors, clinical evidence

      Donors were not screened by a review of relevant medical records for clinical evidence of communicable disease agents and diseases.

    5. Plasma diluted test specimens

      Donors were not determined to be ineligible whose communicable disease test specimens were plasma diluted, and an appropriate algorithm was not used to determine that the plasma dilution did not affect the test results.

    Show 4 more citations
    1. Deviations--evaluation, cause, corrective action

      Investigation of deviations related to core CGTP requirements did not include efforts to determine the cause of the deviation and corrective action(s) to address the deviation and prevent recurrence.

    2. Availability of procedures

      Procedures for core CTGP requirements were not available to personnel in the area where operations are performed, or in a nearby area when such availability is impractical.

    3. Completion of eligibility determination

      HCT/Ps were not kept in quarantine until completion of the donor-eligibility determination.

    4. Ensurane of compliance

      You did not ensure that establishment(s) that by contract, agreement or arrangement, perform manufacturing steps for you were in compliance with applicable CGTP requirements after information became available that suggested the establishment was no longer in compliance.

  6. VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process validation procedures

      Procedures to validate and approve processes that cannot be fully verified by inspection and tests were not followed.

    2. Birth mother sample not tested

      A sample from the birth mother of a donor one month of age or younger was not used for testing of communicable disease agents.

    3. Ensurane of compliance

      You did not ensure that establishment(s) that by contract, agreement or arrangement, perform manufacturing steps for you were in compliance with applicable CGTP requirements prior to the initiation of the contract, agreement of arrangement and applicable CGTP requirements after information became available that suggested the establishment was no longer in compliance.

    4. Documentation elements for activities

      Documentation for activities related to the receipt of HCT/Ps did not include identification of the HCT/P and the establishment that supplied the HCT/P and activities performed and the results of each activity.

    5. Records incomplete

      Records did not identify the person performing the work and were not detailed as necessary to provide a complete history of work performed.

    Show 5 more citations
    1. Determination based on screening and testing

      HCT/P donors were not determined to be eligible based on the results of donor screening and testing.

    2. Responsible person to determine, document

      The eligibility of an HCT/P donor was not determined and documented by a responsible person, based on results of donor screening and donor testing.

    3. Improper release prevention--storage, I.D.

      HCT/Ps from ineligible donors are not stored in a manner to prevent improper release.

    4. Procedures for all other requirements

      Procedures for all steps performed in the testing and screening of donor eligibility of HCT/Ps were not established and followed.

    5. Aseptic processing---cleaning, disinfecting

      Environmental controls do not provide for adequate disinfecting of equipment to ensure aseptic processing.

Regulations FDA cited here

The 17 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

  • 21 CFR 1271.80What are the general requirements for donor testing?4 of 4
  • 21 CFR 1271.65How do I store an HCT/P from a donor determined to be ineligible, and what uses of the HCT/P are not prohibited?2 of 4
  • 21 CFR 1271.150Current good tissue practice requirements2 of 4
  • 21 CFR 1271.160Establishment and maintenance of a quality program2 of 4
  • 21 CFR 1271.230Process validation2 of 4
  • 21 CFR 1271.265Receipt, predistribution shipment, and distribution of an HCT/P2 of 4
  • 21 CFR 1271.47What procedures must I establish and maintain?1 of 4
  • 21 CFR 1271.50How do I determine whether a donor is eligible?1 of 4
Show all 17

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3004195871 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3004195871. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.