Peninsula Medical Products, LLC
31330 Schoolcraft Rd Ste 200, Livonia, Michigan, United States
FEI 3003991756
FDA has 4 inspections on file for Peninsula Medical Products, LLC in Livonia, Michigan (FEI 3003991756), from 2010 to 2016. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.22 and 21 CFR 820.80(b).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2010–2016
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Quality Audit/Reaudit - documentation
The dates of quality audits have not been documented.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptable supplier records, inadequate records
Records of acceptable suppliers and contractors have not been adequately established.
Document review, approval by designated individual
Documents were not reviewed and not approved by designated individual(s) prior to issuance .
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training - Lack of or inadequate procedures
Procedures for identifying training needs were not established.
Design history file
The design history file was not established or maintained.
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals.
Design control - no procedures
Procedures to control the design process of the device were not established and documented.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not established.
Show 4 more citations
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established.
Supplier notification of changes
There is no clear agreement from the suppliers and contractors that they will notify you of changes in the product or service.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not established.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 4
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
Show all 10
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 820.198Complaint files1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003991756 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003991756. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.