FDA has 3 inspections on file for 3CPM Company., INC.- EndoSure, Inc. in Sparks Glenco, Maryland (FEI 3002900388), from 2009 to 2015. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Nov 6, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.30(g), 21 CFR 820.100(b) and 21 CFR 820.184.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2009–2015
- 1 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - software validation not performed
Validation of device software is inadequate.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Acceptable supplier records, inadequate records
Records of acceptable suppliers have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Change records, content
Records of changes did not include a description of the change.
Show 1 more citation
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the verification or validation of corrective actions, implementation of corrective and preventive actions, and submission of information on quality problems and corrective and preventive actions for management review.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization and satisfies the requirements of the quality system regulations.
Lack of or inadequate DHR procedures
Procedures have not been to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Lack of procedures, or not maintained
Document control procedures were not established.
Distribution records
Distribution records do not include the identification and quantity of devices shipped, the date shipped, and control numbers.
Show 6 more citations
Device identification, control numbers
The device history record does not include any device identifications and control numbers used.
Evaluation of suppliers, contractors, etc., requirements
Adequate quality requirements that must be met by contractors and consultants were not established.
Design control - no procedures
Procedures to control the design process of the device were not established.
Acceptance status
The acceptance status of product was not clearly identified.
DMR device specifications
The device master record does not include or refer to the location of specifications.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.50Purchasing controls3 of 3
- 21 CFR 820.40[Reserved]2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.181Device master record2 of 3
- 21 CFR 820.184Device history record2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
Show all 11
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.86Acceptance status1 of 3
- 21 CFR 820.160Distribution1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002900388 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002900388. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.