3CPM Company., INC.- EndoSure, Inc.

53 Loveton Cir Ste 203, Sparks Glenco, Maryland, United States

FEI 3002900388

FDA has 3 inspections on file for 3CPM Company., INC.- EndoSure, Inc. in Sparks Glenco, Maryland (FEI 3002900388), from 2009 to 2015. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Nov 6, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.30(g), 21 CFR 820.100(b) and 21 CFR 820.184.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2009–2015
1 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - software validation not performed

      Validation of device software is inadequate.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    4. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    5. Design verification - Lack of or inadequate procedures

      Procedures for design verification have not been adequately established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

    2. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    3. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    4. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    5. Change records, content

      Records of changes did not include a description of the change.

    Show 1 more citation
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, the verification or validation of corrective actions, implementation of corrective and preventive actions, and submission of information on quality problems and corrective and preventive actions for management review.

    2. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization and satisfies the requirements of the quality system regulations.

    3. Lack of or inadequate DHR procedures

      Procedures have not been to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.

    4. Lack of procedures, or not maintained

      Document control procedures were not established.

    5. Distribution records

      Distribution records do not include the identification and quantity of devices shipped, the date shipped, and control numbers.

    Show 6 more citations
    1. Device identification, control numbers

      The device history record does not include any device identifications and control numbers used.

    2. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by contractors and consultants were not established.

    3. Design control - no procedures

      Procedures to control the design process of the device were not established.

    4. Acceptance status

      The acceptance status of product was not clearly identified.

    5. DMR device specifications

      The device master record does not include or refer to the location of specifications.

    6. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 11

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002900388 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002900388. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.