FDA has 4 inspections on file for Electro-Diagnostic Imaging, Inc. in Milpitas, California (FEI 3002900377), from 2010 to 2017. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.30(g).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
4, 2010–2017
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Personnel

      Personnel do not have the necessary education, background and training to perform their jobs.

    2. Design input - documentation

      Design input requirements were not documented.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Design history file

      The design history file was not established and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Design output - documentation

      Design output was not documented before release.

    Show 1 more citation
    1. Design validation - software validation documentation

      Results of the validation of the device software were not adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    3. Design validation - software validation not performed

      Validation of device software is incomplete.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Retention period

      Required records are not retained for the design and expected life of the device.

    2. Documentation of software validation

      Software validation activities for computers or automated data processing systems used as part of production and the quality system have not been fully documented.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective identification and evaluation of events that may be subject to medical device reporting requirements.

    4. Processing MDRs {see also 803, 804}

      Complaint handling procedures have not been completed to ensure that all complaints are evaluated to determine whether the complaint should be filed as a Medical Device Report.

    5. Identification of actions needed

      Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.

    Show 2 more citations
    1. Justification for not reporting

      A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups, and reviews by a designated person was not included in the record.

    2. Uniform and timely processing

      Complaint handling procedures have not been established to ensure that all complaints are processed in a uniform and timely manner.

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002900377 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002900377. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.