Skip to content

Fareva Unterach GmbH

Mondseestrase 11, Unterach Am Attersee, Austria

FEI 3002829723DUNS 300637281

FDA has 7 inspections on file for Fareva Unterach GmbH in Unterach Am Attersee, Austria (FEI 3002829723), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.

The most recent OAI inspection ended on Oct 23, 2012. FDA issued a warning letter on (CDER). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 211.160(b), 21 CFR 211.165(e) and 21 CFR 314.81(b)(1)(ii).

As of Nov 21, 2026, the site is registered with FDA as a drug establishment for manufacturing, packing, labeling, analysis, sterilization, relabeling and repacking, through Dec 31, 2027. Its registrant is FAREVA Unterach GmbH (D-U-N-S 300637281).

Sources:FDA firm profile ↗FDA drug establishment registrations ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2010–2022
1 OAI · 5 VAI · 1 NAI
Inspections with 483 observations shown
3 of 7
Form 483s posted by FDA
None held for this FEI
Import refusals
1 refusal
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Drug establishment registration
Through

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
  3. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate sampling plans designed to assure that components conform to appropriate standards of identity, strength, quality and purity.

    2. Written program not followed

      The written stability testing program is not followed. Electronic records are used, but they do not meet requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.

  4. NAINo objectionable conditions that need follow-upDrugs · Drug Quality AssuranceNone
  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Certificates of Testing (Containers, Closures)

      Certificates of testing of closures are accepted in lieu of testing without establishing the reliability of the supplier's test results through appropriate validation of the test results at appropriate intervals.

    2. Calibration/Inspection/Checking not done

      Routine calibration of automatic and electronic equipment is not performed according to a written program designed to assure proper performance.

    3. Scientifically sound laboratory controls

      Laboratory controls do not include the establishment of scientifically sound and appropriate specifications, sampling plans, and test procedures designed to assure that drug products conform to appropriate standards of identity, strength, quality and purity.

    4. SOPs not followed / documented

      Written production and process control procedures are not followed in the execution of production and process control functions.

    5. Failure to meet specifications

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning a failure of one or more distributed batches of a drug to meet the specifications established for it in the application.

    Show 3 more citations
    1. Test methods

      The accuracy and specificity of test methods have not been established and documented.

    2. Equipment Design, Size and Location

      Equipment used in the manufacture, processing, packing or holding of drug products is not of appropriate design and suitably located to facilitate operations for its intended use.

    3. Procedures not in writing, fully followed

      The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Validation lacking for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile do not include adequate validation of the sterilization process.

    2. Product flow through building is inadequate

      The flow of components though the building is not designed to prevent contamination.

    3. Equipment to control conditions

      Aseptic processing areas are deficient regarding systems for maintaining any equipment used to control the aseptic conditions.

    4. Environmental Monitoring System

      Aseptic processing areas are deficient regarding the system for monitoring environmental conditions.

    5. Contamination, chemical or physical change, deterioration

      An NDA-Field Alert Report was not submitted within three working days of receipt of information concerning significant chemical, physical, or other change or deterioration in a distributed drug product.

    Show 6 more citations
    1. Written record of investigation incomplete

      Written records of investigations into unexplained discrepancies and the failure of a batch or any of its components to meet specifications do not always include the conclusions and follow-up.

    2. Cleaning / Sanitizing / Maintenance

      Equipment and utensils are not cleaned, maintained, and sanitized at appropriate intervals to prevent malfunctions and contamination that would alter the safety, identity, strength, quality or purity of the drug product.

    3. Cleaning System

      Aseptic processing areas are deficient regarding the system for cleaning and disinfecting the equipment to produce aseptic conditions.

    4. Procedures for sterile drug products

      Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not followed.

    5. In-process and laboratory control results

      Batch production and control records do not include laboratory control results for each batch of drug product produced.

    6. Test methods

      The accuracy, sensitivity, specificity, and reproducibility of test methods have not been established.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 7 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDER · DrugsFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002829723, one row per refused product line.

1 refusal on 2025-12-01 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis

FDA firm profile
  • 2025-12-01

    METHOTREXATE (ANTI-NEOPLASTIC)62IDA31

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG

    Division of Southeast ImportsAEK-7885543-2/10002/1

FDA drug establishment registration

FDA registration search
Registered operations
Manufacturing, packing, labeling, analysis, sterilization, relabeling, repacking
Registration runs through
Dec 31, 2027
Registrant
FAREVA Unterach GmbHD-U-N-S 300637281

From FDA’s Drug Establishments Current Registration Site, as of Nov 21, 2026. A registration says what a site told FDA it does. It is not an inspection result. FDA’s registration search looks firms up by name.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002829723 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002829723. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.