Skip to content

Shirakawa Olympus Co., Ltd.

Azaokamiyama 3-1, Oazaodakura; Nishigo-Mura, Nishishirakawa-Gun, Japan

FEI 3002808148

FDA has 4 inspections on file for Shirakawa Olympus Co., Ltd. in Nishishirakawa-Gun, Japan (FEI 3002808148), from 2012 to 2025. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

6 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.30(g).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
4, 2012–2025
1 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
2 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest May 1, 2026
Import refusals
6 refusals
Import alerts
89-04Red List
Matched on FDA’s firm name and street address

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not where necessary.

    Show 1 more citation
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on May 1, 2026.

  • Inspection ending Nov 14, 2025OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 21, 2026. FDAgent holds no copy that FDA posted.

    1. Jan 15, 2026FDA-FOIA-2026-513Form 483· 1 of 4 sitesRecords released
  • Inspection date not given

    1 request. None of FDA’s log entries for these requests say records were released.

    1. May 1, 2026FDA-FOIA-2026-4360EIR· 1 of 8 sitesClosed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 6 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 4 inspections, with how many of them cited each.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002808148, one row per refused product line.

6 refusals from 2018-10-23 to 2020-04-09 · charged FRNMFGREG on 4, REGISTERED on 4 · no FDA sample analysis

FDA firm profile
  • 2020-04-09

    ULTRASOUND BRONCHOSCOPE77PSV

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6396684-9/11/4

  • 2020-04-09

    TRANSDUCER, ULTRASONIC, DIAGNOSTIC90ITX

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6396684-9/11/3

  • 2020-04-09

    COLONOSCOPE, GASTRO-UROLOGY78FDF

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6396684-9/11/2

  • 2020-04-09

    TRANSDUCER, ULTRASONIC, DIAGNOSTIC90ITX

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6396684-9/11/1

  • 2018-10-23

    TAPE, TELEVISION & VIDEO, CLOSED-CIRCUIT, USED DURING ENDOSCOPIC PROCEDURE78FET

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Northeast Imports510-8375328-8/91/1

Show 1 more
  • 2018-10-23

    TAPE, TELEVISION & VIDEO, CLOSED-CIRCUIT, USED DURING ENDOSCOPIC PROCEDURE78FET

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Northeast Imports510-8375328-8/21/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2026-10-19

    FDA’s entry
    Shirakawa Olympus Co., Ltd. · Azaokamiyama 3-1, Oazaodakura; Nishigo-Mura, Nishishirakawa-Gun, Fukushima JAPAN
    FDA’s inspection record for FEI 3002808148
    Shirakawa Olympus Co., Ltd. · Azaokamiyama 3-1, Oazaodakura; Nishigo-Mura, Nishishirakawa-Gun, Japan

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    6 products

    • 77 E - - OBNasopharyngoscope (Flexible Or Rigid)
    • 77 E - - OQBronchoscope (Flexible Or Rigid)
    • 78 F - - DAEnteroscope
    Show 3 more products
    • 78 F - - GBUreteroscope
    • 78 N - - EZSystem, Imaging, Gastrointestinal, Wireless, Capsule
    • 85 H - - ETLaparoscope, Gynecologic (And Accessories)

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002808148 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002808148. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.