Sagami Rubber Industries Co., Ltd.
2-1, Moto-Machi, Atsugi, Japan
FEI 3002808077
FDA has 2 inspections on file for Sagami Rubber Industries Co., Ltd. in Atsugi, Japan (FEI 3002808077), from 2011 to 2015. The latest ended on and was classified VAI, voluntary action indicated. FDA issued a warning letter on (CDRH).
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection: 5 under 21 CFR Part 820, the device quality system rules, and 2 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
7 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2011–2015
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 7 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Show 2 more citations
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been documented in the device history record.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Documentation
Acceptance activities were not adequately maintained as part of the device history record.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Regulations FDA cited here
The 9 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.184Device history record1 of 2
Show all 9
- 21 CFR 820.198Complaint files1 of 2
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002808077, one row per refused product line.
7 refusals from 2017-01-09 to 2022-04-05 · charged NO ENGLISH on 6, NOT LISTED on 2 · no FDA sample analysis
FDA firm profile- 2022-04-05
CONDOM85HIS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Northern Border ImportsSCS-0685482-5/11/1
- 2019-06-27
CONDOM85HIS
NO ENGLISH 502(c); 801(a)(3);Misbranding
Division of West Coast Imports349-1006847-6/221/1m
- 2017-03-27
CONDOM85HIS
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO ENGLISH 502(c); 801(a)(3);Misbranding
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsUPS-2240226-8/61/1A
- 2017-03-09
CONDOM85HIS
NO ENGLISH 502(c); 801(a)(3);Misbranding
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southeast ImportsAEK-3005412-3/11/1
- 2017-02-16
CONDOM85HIS
NO ENGLISH 502(c); 801(a)(3);Misbranding
Cincinnati District OfficeUPS-2053613-3/31/1C
Show 2 more
- 2017-02-16
CONDOM85HIS
NO ENGLISH 502(c); 801(a)(3);Misbranding
Cincinnati District OfficeUPS-2053613-3/31/1B
- 2017-01-09
CONDOM85HIS
NO ENGLISH 502(c); 801(a)(3);Misbranding
Cincinnati District OfficeUPS-1745998-4/71/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002808077 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002808077. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.