Biosensors Interventional Technologies Pte Ltd
36 Jalan Tukang, Singapore, Singapore
FEI 3002806593
FDA has 3 inspections on file for Biosensors Interventional Technologies Pte Ltd in Singapore, Singapore (FEI 3002806593), from 2010 to 2023. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 5 citations from this inspection: 3 under 21 CFR Part 803, the device adverse event reporting rules, and 2 under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 3, 2010–2023
- 2 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Jul 11, 2024
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Project Evaluation: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Product rework adverse effects {see also 820.184}
Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Supplier oversight
The type and extent of control to be exercised over suppliers was not clearly defined.
Show 1 more citation
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Jul 11, 2024. It asked for a Form 483.
Inspection ending Apr 20, 2023OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Jul 11, 2024FDA-FOIA-2024-6084Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 7 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002806593, one row per refused product line.
1 refusal on 2020-04-17 · charged NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2020-04-17
INTRODUCER, CATHETER74DYB
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-6512456-1/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3002806593 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3002806593. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.