Skip to content

Future Diagnostics Solutions B.V.

Nieuweweg 279, Wijchen, Netherlands

FEI 3001857593

FDA has 4 inspections on file for Future Diagnostics Solutions B.V. in Wijchen, Netherlands (FEI 3001857593), from 2011 to 2024. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

2 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-04 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.

Four regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
4, 2011–2024
2 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Aug 18, 2025
Warning letters
None on file
Import refusals
2 refusals
Import alerts
89-04Red List
Matched on FDA’s firm name and street address

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design input - documentation

      Design input requirements were not adequately documented.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    4. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design review - documentation

      The design review results, including the individual(s) performing the review, were not documented in the design history file.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Aug 18, 2025. Each one asked for a Form 483.

  • Inspection ending Jul 11, 2024OAI

    1 request. In the inspection history above.

    1. Oct 25, 2024FDA-FOIA-2024-9511Form 483Open
  • Inspection ending Jan 19, 2023OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Dec 29, 2025. FDAgent holds no copy that FDA posted.

    1. Aug 18, 2025FDA-FOIA-2025-9964Form 483· 1 of 2 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 7 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001857593, one row per refused product line.

2 refusals from 2017-12-29 to 2025-03-04 · charged DEVICEGMPS on 1, NOT LISTED on 1 · no FDA sample analysis

FDA firm profile
  • 2025-03-04

    IMMUNOASSAY METHOD, TROPONIN SUBUNIT75MMI

    DEVICEGMPS 501(h), 801(a)(1); ADULTERATION

    Division of Southeast Imports799-6014432-8/11/1

  • 2017-12-29

    INDICATOR, BIOLOGICAL STERILIZATION PROCESS80FRC

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southeast Imports799-3594023-6/11/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-04Red ListDetention Without Physical Examination of Devices from Firms that Have not met Device Quality System RequirementsImport alert pageAlert on fda.gov

    Date published 2024-11-21

    FDA’s entry
    Future Diagnostics Solutions B.V. · Nieuweweg 279, Wijchen, Gelderland NETHERLANDS
    FDA’s inspection record for FEI 3001857593
    Future Diagnostics Solutions B.V. · Nieuweweg 279, Wijchen, Netherlands

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    1 product

    • 75 - - - --Chemistry

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001857593 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001857593. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.