P.T. Sankei Medical Industries
Jl. Mordern Inustri Iii, Komp. Bpsp A- 12, Serang, Indonesia
FEI 3001576493
FDA has 2 inspections on file for P.T. Sankei Medical Industries in Serang, Indonesia (FEI 3001576493), from 2016 to 2023. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 8 citations from this inspection: 7 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
5 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2016–2023
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 1 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 5 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Show 3 more citations
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.181Device master record1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesLetter on fda.govFDA record
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001576493, one row per refused product line.
5 refusals on 2024-04-01 · charged NOT LISTED on 5, DIRECTIONS on 5 · no FDA sample analysis
FDA firm profile- 2024-04-01
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
Charge 3904
Division of West Coast ImportsEI5-0070787-9/51/1
- 2024-04-01
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Charge 3904
Division of West Coast ImportsEI5-0070787-9/41/1
- 2024-04-01
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
Charge 3904
Division of West Coast ImportsEI5-0070787-9/31/1
- 2024-04-01
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Charge 3904
Division of West Coast ImportsEI5-0070787-9/21/1
- 2024-04-01
SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN
NOT LISTED 502(o), 801(a)(3); MISBRANDING
DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Charge 3904
Division of West Coast ImportsEI5-0070787-9/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3001576493 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3001576493. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.