Skip to content

P.T. Sankei Medical Industries

Jl. Mordern Inustri Iii, Komp. Bpsp A- 12, Serang, Indonesia

FEI 3001576493

FDA has 2 inspections on file for P.T. Sankei Medical Industries in Serang, Indonesia (FEI 3001576493), from 2016 to 2023. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 8 citations from this inspection: 7 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

5 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
2, 2016–2023
1 OAI · 1 VAI · 0 NAI
Inspections with 483 observations shown
1 of 2
Form 483s posted by FDA
None held for this FEI
Import refusals
5 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    5. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    Show 3 more citations
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Design control - no procedures

      Procedures for design control have not been established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published

Regulations FDA cited here

The 8 sections of 21 CFR in FDA’s published citations for 1 of this facility’s 2 inspections.

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3001576493, one row per refused product line.

5 refusals on 2024-04-01 · charged NOT LISTED on 5, DIRECTIONS on 5 · no FDA sample analysis

FDA firm profile
  • 2024-04-01

    SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    Charge 3904

    Division of West Coast ImportsEI5-0070787-9/51/1

  • 2024-04-01

    SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Charge 3904

    Division of West Coast ImportsEI5-0070787-9/41/1

  • 2024-04-01

    SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    Charge 3904

    Division of West Coast ImportsEI5-0070787-9/31/1

  • 2024-04-01

    SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Charge 3904

    Division of West Coast ImportsEI5-0070787-9/21/1

  • 2024-04-01

    SYSTEM, MEASUREMENT, BLOOD-PRESSURE, NON-INVASIVE74DXN

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    DIRECTIONS 502(f)(1), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Charge 3904

    Division of West Coast ImportsEI5-0070787-9/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3001576493 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3001576493. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.