Adamson Analytical Laboratories, Inc.
220 Crouse Dr, Corona, California, United States
FEI 3000204955
FDA has 6 inspections on file for Adamson Analytical Laboratories, Inc. in Corona, California (FEI 3000204955), from 2013 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 211, the manufacturing rules for finished drugs. It is listed below. One of the three FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Oct 4, 2022. FDA has issued 2 warning letters under this FEI, the latest on (CDER · Division of Pharmaceutical Quality Operations IV). No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2013–2024
- 3 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest FY2023
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Establishment of calibration procedures
Procedures describing the calibration of instruments, apparatus, gauges and recording devices are deficiently written or followed.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training--operations, GMPs, written procedures
Employees are not given training in the particular operations they perform as part of their function.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Test methods
The accuracy, sensitivity, specificity and reproducibility of test methods have not been established and documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality AssuranceCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Laboratory equipment calibration records
Laboratory records do not include complete records of the periodic calibration of laboratory instruments, apparatus, gauges and recording devices.
Instruments, apparatus, et. al. not meeting specs
The use of apparatus not meeting established specifications was observed.
Approve or reject procedures or specs
The quality control unit lacks responsibility to approve and reject all procedures or specifications impacting on the identity, strength, quality and purity of drug products.
Calibration - at intervals, written program, remedial action
The calibration of instruments and apparatus is not done at suitable intervals in accordance with an established written program and with provisions for remedial action in the event accuracy and/or precision limits are not met.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDrugs · Drug Quality Assurance12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Cleaning/maintenance records not kept
Records are not kept for the maintenance, cleaning, sanitizing, and inspection of equipment.
Testing and standardization of standards et. al.
Laboratory records do not include complete records of any testing and standardization of laboratory reference standards, reagents, and standard solutions.
Complete test data included in records
Laboratory records do not include complete data derived from all tests, examinations and assay necessary to assure compliance with established specifications and standards.
Computer control of master formula records
Appropriate controls are not exercised over computers or related systems to assure that changes in master production and control records or other records are instituted only by authorized personnel.
Approve or reject procedures or specs
The quality control unit lacks responsibility to approve and reject all procedures or specifications impacting on the identity, strength, quality, and purity of drug products.
Show 7 more citations
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not cleaned, maintained, and sanitized at appropriate intervals to prevent malfunctions and contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Defined areas of adequate size for operations
The separate or defined areas necessary to prevent contamination or mix-ups are deficient.
Calibration/Inspection/Checking not done
Routine inspection and checking of automatic, mechanical, and electronic equipment is not performed according to a written program designed to assure proper performance.
Training--operations, GMPs, written procedures
Employees are not given training in the particular operations they perform as part of their function, current good manufacturing practices, and written procedures required by current good manufacturing practice regulations.
Approval and review of procedures
Written procedures are not drafted, reviewed and approved by the appropriate organizational units and reviewed and approved by the quality control unit.
Buildings not maintained in good state of repair
Buildings used in the manufacturing, processing, packing, and holding of a drug product are not maintained in a good state of repair.
Calibration - at intervals, written program, remedial action
The calibration of instruments and recording devices is not done at suitable intervals in accordance with an established written program and with provisions for remedial action in the event accuracy and/or precision limits are not met.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality AssuranceCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Sep 12, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 4, 2024FDA-FOIA-2024-8758Form 483· 1 of 5 sitesClosed
Inspection ending Oct 4, 2022OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-2871Form 483Closed
Inspection date not given
1 request. None of FDA’s log entries for these requests say records were released.
- Mar 27, 2025FDA-FOIA-2025-2817Form 483 + EIRClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 211.22Responsibilities of quality control unit3 of 4
- 21 CFR 211.160General requirements3 of 4
- 21 CFR 211.25Personnel qualifications2 of 4
- 21 CFR 211.68Automatic, mechanical, and electronic equipment2 of 4
- 21 CFR 211.194Laboratory records2 of 4
- 21 CFR 211.42Design and construction features1 of 4
- 21 CFR 211.58Maintenance1 of 4
- 21 CFR 211.67Equipment cleaning and maintenance1 of 4
Show all 10
- 21 CFR 211.100Written procedures; deviations1 of 4
- 21 CFR 211.165Testing and release for distribution1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDER · Division of Pharmaceutical Quality Operations IV · DrugsLetter on fda.govFDA record
- Warning letterFormal notice of significant violationsCDER · DrugsFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000204955 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000204955. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.