Life Technologies Corporation
9099 N. Deerbrook Trail, Brown Deer, Wisconsin, United States
FEI 2244574
FDA has 4 inspections on file for Life Technologies Corporation in Brown Deer, Wisconsin (FEI 2244574), from 2008 to 2015. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 6 citations from this inspection: 2 under 21 CFR Part 803, the device adverse event reporting rules, 2 under 21 CFR Part 806, the device correction and removal reporting rules, and 2 under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2008–2015
- 0 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Device number, identification
The total number of devices subject to correction or removal actions was not reported.
Domestic & foreign consignees
The names, addresses, and telephone numbers of all domestic and foreign consignees of devices subject to correction or removal actions and number of devices distributed to each consignee were not reported.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Submission of Supplement to Individual Report Within One Month
A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design review - documentation
The design review results, including the date and the individual(s) performing the review, were not documented in the design history file.
Device number, identification
The total number of devices subject to correction or removal actions was not reported.
Domestic & foreign consignees
The names, addresses, and telephone numbers of all domestic and foreign consignees of devices subject to correction or removal actions and number of devices distributed to each consignee were not reported.
Product rework adverse effects {see also 820.184}
Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 3 more citations
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Resources
Adequate resources have not been provided for performing management activities, work, and assessment activities.
Training
Employees have not been adequately trained.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not complete and implemented.
Verify, validate change {see also 820.100(a)}
The procedures addressing verification or validation of corrective and preventive actions were not documented, complete, and implemented.
Regulations FDA cited here
The 8 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?2 of 3
- 21 CFR 806.10Reports of corrections and removals2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
- 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
Other FEIs recorded as Life Technologies Corporation
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Life Technologies CorporationFEI 3009976420Pleasanton, California, United States
Last inspected 2026-08-13NAI
Life Technologies CorporationFEI 3007287152Carlsbad, California, United States
Last inspected 2026-07-23NAI
Life Technologies CorporationFEI 3015529000Guilford, Connecticut, United States
Last inspected 2025-05-21NAI
Life Technologies CorporationFEI 3003335080Frederick, Maryland, United States
Last inspected 2021-06-14VAI
Life Technologies CorporationFEI 1317268Grand Island, New York, United States
Last inspected 2015-10-02NAI
Life Technologies CorporationFEI 2917132Camarillo, California, United States
Last inspected 2009-08-19VAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2244574 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2244574. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.